Vetscan Feline Leukemia Virus Antigen-Feline Immunodeficiency Virus Antibody Test (Canada)
This treatment applies to the following species:RAPID FeLV-FIV TEST
For the qualitative detection of feline leukemia virus antigen and feline immunodeficiency virus antibodies in feline anticoagulated whole blood, serum, or plasma
Intended Use
The Vetscan® Feline Leukemia Virus Antigen-Feline Immunodeficiency Virus Antibody Rapid Test is a visual rapid test for the qualitative detection of feline leukemia virus antigen and feline immunodeficiency virus antibody in feline anticoagulated whole blood, serum, or plasma. This test is for veterinary use only.
INSTRUCTIONS FOR USE
● Samples must be at room temperature (15-27 °C/59-80 °F), before running the assay - DO NOT HEAT.
● Previously frozen or older samples must be centrifuged before use.
● Serum or Plasma, either fresh, previously frozen or stored at 2-7 °C (36-45 °F), may be used in this test. Serum or plasma may be stored for up to 7 days at 2-7 °C. For longer storage, sample should be frozen at -20 °C (-4 °F) or colder.
● Whole Blood Must be anticoagulated (e.g. EDTA, heparin) and may be used either fresh or after refrigeration at 2-7 °C (36-45 °F) for up to 5 days.
● Excessive hemolysis may obscure the results.
● EDTA or heparin in plasma will not affect the results.
PRECAUTIONS AND WARNINGS
● Important: Do not remove Test Device from the pouch until ready for use.
● For veterinary use only.
● Do not use components after expiration date.
● The Test Device must be used as soon as possible after removing from pouch.
● The Test Device should be in a horizontal position on a flat surface while the test is performed.
● Use a separate Transfer Pipette for each test.
● The Chase Buffer is not interchangeable from serial (lot) to serial (lot).
● Contains sodium azide as a preservative.
● Handle samples as biohazardous materials.
● Dispose of contents/container in accordance with local/regional/national/international regulations.
STORAGE
● Test Devices and Chase Buffer must be stored at room temperature 15-27 °C (59-80 °F).
● Test Devices and Chase Buffer are stable until the expiration date when stored at 15-27 °C (59-80 °F).
KIT COMPONENTS
● Test Devices
● Chase Buffer Bottle (6 mL)
● Transfer Pipettes
● Instructions for Use
TEST PROCEDURE
1. Remove the Test Device from the protective pouch and place on a flat surface. Label the Test Device with the subject I.D. or control identification. Only use the provided white space for this purpose. Do not cover up the test name.
2. Gently mix the sample by inverting.
3. Using a provided Transfer Pipette, hold the pipette vertically and dispense one drop of sample (anticoagulated whole blood, serum, or plasma) into sample well.
4. Holding the Chase Buffer Bottle vertically, promptly add 2 drops of the chase buffer to the sample well.
5. Read the results within 10 minutes. Do not read results after 10 minutes. Colored lines which appear after 10 minutes are not diagnostic and should be ignored.

INTERPRETATION OF TEST RESULTS
Positive Results
A test is positive if a colored line appears under the FeLV (L) or FIV (I) areas and a line appears under the Control line (C) area. Any intensity of the Test line L or I areas should be considered positive. Colored lines may be lighter or darker than each other.
Negative Results
The test is negative if only one line appears at the Control line (C) area.
Invalid Results
The test is invalid if no colored line appears at the Control line (C) area even if a colored line appears in the Test line L or I areas. If a colored line does not appear in the C area, the test is invalid and should be repeated.
Manufacturer
SA Scientific, 4919 Golden Quail, San Antonio, TX 78240 USA
VLN/PCN 373/502A.01
1-210-699-8800
Distributor
Zoetis Inc., 333 Portage St., Kalamazoo, MI 49007 USA
US 1-888-963-8471
Canada 1-800-461-0917
Imported by
Zoetis Belgium S.A., Rue Laid Burniat 1, 1348 Louvain-La-Neuve Belgium
www.zoetis.com

51285702
Presentation: 10 and 25 test kits.
CPN: 1198588.1
16,740 TRANS-CANADA HIGHWAY, KIRKLAND, QC, H9H 4M7
Order Desk: | 800-663-8888 | |
Technical Services Canada: | 800-461-0917 | |
Technical Services USA: | 800-366-5288 | |
Website: | www.zoetis.ca |
![]() |
THIS SERVICE AND DATA ARE PROVIDED "AS IS". Animalytix assumes no liability, and each user assumes full risk, responsibility, and liability, related to its use of the Animalytix service and data. See the Terms of Use for further details. |
Copyright © 2025 Animalytix LLC. Updated: 2025-05-29
Elsewhere on our site...
Wegovy
Wegovy (semaglutide) an FDA-approved weekly injection for weight loss and to reduce heart risks ...
Rybelsus
Rybelsus tablets are used to improve blood sugar control in adults with type 2 diabetes, and may ...
Ozempic
Learn about Ozempic (semaglutide) for type 2 diabetes treatment, weight management, cardiovascular ...
Zepbound
Zepbound (tirzepatide) is an FDA-approved weekly injection for weight loss and obstructive sleep ...
Trulicity
Trulicity is an injectable diabetes medicine that is used together with diet and exercise to ...
Lantus
Lantus is a long acting form of insulin used to treat type 1 or type 2 diabetes. Learn about side ...
Tresiba
Tresiba (insulin degludec) is used to treat diabetes mellitus. Includes Tresiba side effects ...
Phentermine
Phentermine is an appetite suppressant used together with diet and exercise to treat obesity. Learn ...
Semaglutide
Semaglutide is a GLP-1 agonist used for weight loss, type 2 diabetes, and reducing cardiovascular ...