Skip to main content

Patch-Wearable Cardioverter-Defibrillator Safe and Effective

Medically reviewed by Carmen Pope, BPharm. Last updated on Aug 2, 2024.

By Lori Solomon HealthDay Reporter

FRIDAY, Aug. 2, 2024 -- A novel, water-resistant patch-wearable cardioverter-defibrillator (P-WCD) is safe and effective for patients at risk for sudden cardiac arrest, according to a study published in the Aug. 6 issue of the Journal of the American College of Cardiology.

John Hummel, M.D., from The Ohio State University in Columbus, and colleagues assessed the safety and clinical effectiveness of a novel P-WCD. The analysis included 290 patients at risk for sudden cardiac arrest due to ventricular tachycardia/ventricular fibrillation who were not candidates for or refused an implantable defibrillator.

The researchers found that the clinically significant cutaneous adverse device effect rate was 2.30 percent, with no severe adverse effects. There were no device-related deaths or serious adverse events reported. The inappropriate shock rate was 0.36 per 100 patient-months. Nine patients received 11 shocks, of which nine shocks were adjudicated to be appropriate. Eight of nine appropriate shocks were successful with a single shock. Median wear time compliance was 23.5 hours per day.

"There were no patient deaths or missed episodes requiring external rescue, and high patient compliance enabled a high number of successful life-saving conversions," the authors write.

Several authors disclosed ties to medical device companies, including Element Science, which manufactures the studied P-WCD and funded the trial.

Abstract/Full Text

Editorial (subscription or payment may be required)

Disclaimer: Statistical data in medical articles provide general trends and do not pertain to individuals. Individual factors can vary greatly. Always seek personalized medical advice for individual healthcare decisions.

© 2025 HealthDay. All rights reserved.

Read this next

Real-Time AI-Driven Decision Support Aids Catheter Ablation

WEDNESDAY, May 7, 2025 -- An artificial intelligence (AI)-driven model can significantly improve procedural safety in cardiac electrophysiology with real-time decision support...

Study IDs Nonclinical, Modifiable Risk Factors Tied to Sudden Cardiac Arrest

THURSDAY, May 1, 2025 -- Modifiable, nonclinical risk factors such as lifestyle modifications and physical measures may reduce the risk for sudden cardiac arrest (SCA), according...

Faster Walking Pace Cuts Risk for Cardiac Arrhythmias

FRIDAY, April 25, 2025 -- Average and brisk walking pace are associated with a decreased risk for cardiac arrhythmias, according to a study published online April 15 in...