Zoll LifeVest 4000 Wearable Cardioverter Defibrillator: FDA Safety Communication - Potential Lack of Treatment (Shock) Delivery Due to Device Failure

Audience: Risk Manager, Cardiology, Nursing

[Posted 01/17/2018]

ISSUE: FDA is providing information and recommendations regarding the Zoll LifeVest 4000 due to concerns that the device may fail to deliver treatment to the patient if the device is not replaced soon after displaying "Call for service: Device has a problem that may require service. Call ZOLL for service, Message Code 102." Failure to contact Zoll and immediately replace the device after Message Code 102 appears on the device screen may result in serious patient harm or death of the patient because the device may fail to deliver therapy appropriately when needed.

Ideally, the LifeVest monitors the patient's heart, delivering a "treatment shock" as needed to restore the patient's heartbeat to a normal rhythm. However, FDA is aware that in certain cases the LifeVest 4000 may not be able to deliver a life-saving treatment shock to a patient due to a fault that prevents the device from charging its high-energy capacitors. During normal operation, this fault prompts the LifeVest to display "Call for service: Device has a problem that may require service. Call Zoll for service, Message Code 102" on the device screen. The "Message Code 102" alert does not explicitly indicate to the patient that the device cannot be used and that the patient should call Zoll immediately.

To date, FDA is aware of one patient death due to the LifeVest's failure to deliver treatment as expected after Message Code 102 was displayed. According to Zoll, as of November 14, 2017, there were a total of 33,670 devices distributed, with 24,975 devices distributed in the U.S. About 0.1% of the total devices distributed have displayed Message Code 102, which should only be remedied by immediate replacement of the device.

The FDA will continue to work with Zoll to monitor LifeVest 4000 devices for any adverse events related to Message Code 102 or a failure to deliver treatment. The FDA will also continue to work with Zoll to identify a permanent solution to the Message Code 102 issue, and will keep the public informed as new information becomes available.

BACKGROUND: The Zoll LifeVest 4000 is a wearable defibrillator used to treat life-threatening abnormal heartbeats (arrhythmias) in adults and children who are at risk for sudden cardiac arrest and are not candidates for, or refuse, an implantable defibrillator. The LifeVest continuously monitors the patient's heart and, if a fast, life-threatening heart rhythm is detected, the device delivers a treatment (i.e. a shock) to restore the patient to a normal heart rhythm.

RECOMMENDATION: For Health care providers -

For Patients and Caregivers:

Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information and Adverse Event Reporting Program:

[01/17/2018 - Safety Communication - FDA]