FLOW-i Anesthesia Systems by Maquet: Class 1 Recall - Dislodged Patient Cassette May

[Posted 07/01/2015]

AUDIENCE:Risk Manager

ISSUE: FDA is providing notification of a Class 1 recall of FLOW-i Anesthesia Systemsby Maquet. The firm has received several complaints where patient cassettes, which are the center of gas flow in the system, have come loose. The patient cassette locking device may accidentally release the patient cassette from its mount when usersperform a change of system. This may cause anesthesia gas to leak and could prevent the ventilator from providing breathing support if not corrected immediately.

The firm has received 10 reports this device has malfunctioned; no injuries or deathshave been reported.

BACKGROUND: The Flow-i Anesthesia System administers anesthesia while providing ventilation to patients with no or limited ability to breathe. The system is used in hospitals, for use in a range of patients from neonatal to adult.

RECOMMENDATION: A Maquet Service representative will contact consignees to arrange replacement of the FLOW-i Anesthesia System patient cassette locking device.

For questions regarding this field action, please contact a Maquet Service Representative or Maquet Technical Support at 1-888-627-8383 (Press option 3, followed

by option 1 and then option 1 again), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. EST.


Healthcare professionals and patients are encouraged to report adverse events or side effects related to the use of these products to the FDA's MedWatch Safety Information

and Adverse Event Reporting Program:


[07/01/2015 -Recall Notice - FDA]