Empower

This treatment applies to the following species:
Manufacturer: Metrex

Dual Enzymatic Detergent

Low foaming, dual enzymatic detergent designed for easy one-step pre-cleaning by disintegrating blood and protein from delicate instruments and equipment.

Cleans Fast

In manual, ultrasonic and automatic cleaning systems, EmPower rapidly dissolves and carries away organic soils.

Safe For Instruments And Equipment

EmPower is a neutral pH formula that provides advanced corrosion protection that is safe for rubber, plastic, stainless steel, carbon steel and soft metals.

Safe For Users: Empower Reduces The Need For Manual Handling And Scrubbing, Especially In Hard-to-reach Places Like Channels, Lumens And Instrument Box Locks.

keep Out Of Reach Of Children

Important

Store In A Cool Area.

Risk Phrase(s)

- May cause eye and skin irritation.

- May cause throat, mouth and stomach irritation if ingested.

- Repeated inhalation may result in sensitization and allergic reaction in hypersensitive individuals.

Precautionary Measures

- Avoid contact with strong oxidizing agents and removal of water.

- Avoid spraying or high pressure washing.

- Keep in cool place.

Personal Protective Equipment

- Protective gloves, eye gear, and impervious gown must be worn.

First Aid Measures

Eyes-flush Immediately With Water For 15 Minutes. Seek Medical Attention. Temporary Redness Of Eye May Occur.

skin-wash With Soap And Water.

inhalation-remove To Fresh Air. Consult Physician If Allergic Response Exhibited.

ingestion-do Not Induce Vomiting. Consult Physician.

for Further Information Refer To Metrex Research Corp. Empower Material Safety Data Sheet

directions For Use

Use Fresh Empower® For Each Endoscope Or Set Of Instruments. Discard Diluted Empower Solution After Each Use.

Manual Cleaning

Add 1 oz. of concentrate to one gallon of warm water (68°F-104°F). Soak instruments and equipment immediately after use, until soil is dissolved and removed. Soak for a minimum of 1 minute. Soak longer if matter is dried on, until visibly clean. Immerse instruments completely and aspirate solution into any channels and lumens. Use an instrument brush for light mechanical cleaning if necessary. Following soak period, visually inspect for cleanliness and rinse thoroughly with warm water. Rough dry. After cleaning with EmPower, equipment and instruments must be disinfected/sterilized according to appropriate procedures.

Endoscope Reprocessors

Add 1/2 oz. of concentrate per gallon of water to the reprocessor reservoir, or attach container to the reprocessor's automatic dispensing unit.

Ultrasonic Units

Add 1 oz. of concentrate per gallon of water. Recommended water temperature is 68°F-104°F. A soak time of 2 minutes or more is recommended for ultrasonic cleaners. Process instruments according to manufacturer's directions.

Washer Sterilizers/decontaminators

Attach concentrate container to the unit's automatic dispensing unit.

Note: EmPower is not a sterilant/disinfectant.

Storage And Disposal

Storage

Store in a cool area.

Solution Disposal

For unused solution, flush thoroughly with large quantities of water into sewage disposal system in accordance with federal, state, and local regulations.

For used solution, the waste solution must be characterized by the generator and disposed of in accordance with federal, state and local regulations.

Container Disposal

Do not reuse empty container. Wrap container and put in trash.

For product or technical information, contact Metrex at 800-841-1428 or visit our web site at www.metrex.com.

METREX RESEARCH CORPORATION®, 28210 Wick Road, Romulus, Michigan 48174

Contents:

Reorder No.

 

4 x 1 Gallon / 3.8 Litre bottles per case

10-4100

70-4050-8

gallon dispensing pump

 

 

1 x 5 Gallon (19 L) bottle per case

10-4150

70-4056-9

1 x 15 Gallon (57 L) drum

10-4115

70-4056-9

Nac No.

12660022
METREX RESEARCH™
1717 W. COLLINS AVENUE, ORANGE, CA, 92867
Telephone:   714-516-7400
Fax:   714-516-7904
Website:   www.metrex.com
Email:   info@metrex.com
Every effort has been made to ensure the accuracy of the EmPower information published above. However, it remains the responsibility of the readers to familiarize themselves with the product information contained on the Canadian product label or package insert.
(web1)