Medtronic Neuromodulation SynchroMed EL, SynchroMed II and IsoMed Implantable Infusion Pumps
[Posted 03/21/2008] FDA and Medtronic notified healthcare professionals of the Class 1 recall of Medtronic Neuromodulation Implantable Infusion Pumps, implantable devices for administering drugs to a specific site in the body to treat pain, spasticity and cancer. The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of intrathecal catheters which infuse opioids, baclofen, or chemotherapy drugs into patients. On January 16, 2008, Medtronic sent a letter to doctors who implant these devices and/or provide care to patients with the implanted device. The letter described the problem, patient risks, patient management, recommendations and next steps. Medtronic’s representatives will complete a reply card to document their communication with each doctor regarding this recall, as well as asking the doctors to sign and return a reply card.
[March 21, 2008 - Recall Notice - FDA]
[January, 2008 - Dear Healthcare Professional Letter - Medtronic]
FDA MedWatch Alerts Archive
2007: Jan | Feb | Mar | Apr | May | Jun | Jul | Aug | Sep | Oct | Nov | Dec
2006: Jan | Feb | Mar | Apr | May | Jun | Jul | Aug | Sep | Oct | Nov | Dec
2005: Jan | Feb | Mar | Apr | May | Jun | Jul | Aug | Sep | Oct | Nov | Dec
2004: Jan | Feb | Mar | Apr | May | Jun | Aug | Sep | Oct | Nov | Dec
2003: Jan | Feb | Apr | May | Jul | Aug | Sep | Oct | Nov | Dec
2002: Feb | Mar | Apr | May | Jun | Jul | Sep | Oct | Nov